WARNING
Possible hazards posed by the battery pack
Damaged or swollen battery packs can overheat, catch
fire, or leak. Immediately cease use of CoaguChek Pro II
meters with damaged or swollen battery packs and
under no circumstances recharge them (do not place in
the Handheld Base Unit).
Overheating can cause the battery pack to catch fire or
explode.
Never throw the battery pack or the meters onto a
fire. Do not dismantle, compress, or pierce the
battery pack as this could cause an internal short
circuit that leads to overheating.
Do not place either the battery pack or the
CoaguChek Pro II meter on or in heating appli-
ances, such as a microwave, conventional oven, or
radiator.
Avoid prolonged exposure to direct sunlight, e.g.,
when the meter is docked in the Handheld Base
Unit. Keep this in mind when positioning the Hand-
held Base Unit.
Battery fluid or materials leaking from damaged battery
packs can irritate your skin or cause burns due to high
temperatures.
Avoid contact with leaking battery fluid. In the
event of accidental contact with the skin, rinse with
water. If you get battery fluid in your eye(s), you
should also seek medical attention.
Handle and dispose of battery packs with care.
Extreme temperatures reduce the charging capacity
and usage period of the meter and the battery pack.