82
6 Touch On (the target value will be displayed with a
test result) or Off (the target value will not be dis-
played). Your selection is now highlighted.
7 Use the arrows to to set the allowed percentage
deviation from the target value.
Use the arrows on the left to adjust the number before
the decimal point. Use the arrows on the right to adjust
the number after the decimal point.
8 Touch to save this setting, or touch to exit
this menu without saving any changes. The display
automatically returns to the QC Settings screen.
The Custom Range and Target Value (if set to On) appear
in a line below the control test result in the QC test and
QC memory screens.
In the left hand screen, Display target value is set to Off:
the custom range only is displayed below the control test
result.
In the right hand screen, Display target value is set to On:
both the custom range and the target value are displayed.
The target value (here: 2.9) appears in front of the
custom range.
QC Range
09:15 am
05/29/2015
Deviation from
target value
Display target
value
On Off
11.5 %
QC Test
09:15 am
05/29/2015
3.0
INR
05/29/2015 09:15 am
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 2
(2.6 - 3.2 INR)
16
%Q
(16 - 20 %Q)
QC Test
09:15 am
05/29/2015
3.0
INR
05/29/2015 09:15 am
QC: 999
Code: 184
Op: Operator 1
Par: PT
Level: 2
(2.9, 2.6 - 3.2 INR)
16
%Q
(17, 16 - 20 %Q)