6.11
P = name of parameter
D = digits/decimal points
E = characters for the unit
DD°C = enzyme reference temperature
v = preceding sign, e.g.
*
; <; >
– = space
The arrangement is usually left-adjusted, e.g.
“GGT < 2.80 U/l 25°C”
Notes
If there is no sample/patient identification, the consecutive number
of the sample is transferred in the 3rd line in the automatic data
transfer.
If there is no “additional test information”, line 6 contains only CR LF.
If the result is not provided with a “result-related message”, line 5
consists only of CR LF.
Apart from the test results, the calculated results such as cardiac risk
or LDL cholesterol can be transferred manually (R see Section 3).
The data protocols correspond to the outputs to the printer.
User instructions and error messages are not transferred. They
are only output to the printer.
The following transfer protocol applies when the function key
F8 is used:
Each line corresponds to a record and is composed as follows:
;
Time and date
;
Patient/sample identification
;
Result with parameter, value and dimension
;
Additional test information, compiled by the instrument*
;
Additional parameter information, compiled by the user*
;
CR
LF