6.3
Software
In addition to the software programmed into the instrument for the
performance and evaluation of a test and the software for the auto-
matic checks performed by the instrument, Reflotron Plus also con-
tains software that allows the operator to select various laboratory
specific basic settings and recurring functions via display prompts
and function keys and via an optional keyboard.
Details of the procedure for calling up and performing these functions
are given in the various chapters. These chapters also describe the
individual display messages and how to work with these programs.
Basic instrument settings
The basic instrument settings are carried out by Roche Diagnostics
or your supplier before delivery.
However, if circumstances change, e.g. in the case of a general
change in the system of units to be used for displaying the results, or
if the standby mode is changed or if the Reflotron Plus is connected
to a computer system, it may become necessary to alter the basic
settings.
Switch off the instrument.
Press the two cursor buttons “<” and “>” simultaneously and hold
them down. Then switch on the instrument with the “ON/OFF”
switch, continuing to hold down the cursor buttons “<” and “>” until
the menu appears in the display.
The menu starts with “LANGUAGE”.
Calling up the menu
“Basic settings”
LA N G U A GE : U S A GER U . K .
FR A N C E SP A I N I TAL Y F R E E
General notes