2.21
15 . 0 5 . 01 09 :13 : 0 4
GL U <10 . 0 m g /d l
15 . 0 5 . 01 09 :13 : 0 4
GL U >60 0 m g / dl
15 . 0 5 . 01 09 :13 : 0 4
GG T D I L UT E S A M PLE
15 . 0 5 . 01 09 :13 : 0 4
GG T *2 0 0 U / l 3 7 ° C
15 . 0 5 . 01 09 :13 : 0 4
NO N - L I NE A R R E ACT I O N
Note
If you wish you may change the
units or the enzyme reference
temperature. R See “Individual
settings” in Section 2.
If results are obtained which are
clinically implausible, please
refer to R Section 4. “Cleaning
and Quality Checking”.
– The result is below the lower
limit of the measuring range.
– The test strip was measured
without application of a
sample.
R Repeat the test.
The result is above the upper
limit of the measuring range. If
necessary, dilute the sample.
R See pack insert of the
Reflotron test.
The enzyme activity is too high
or the colour of the sample is
too intense. In either case the
sample must be diluted and the
analysis repeated. R See pack
insert of the Reflotron test.
An asterisk in front of the result
indicates that the reaction
kinetics were slightly non-linear.
Repetition of the test and/or
dilution of the sample is recom-
mended. R See pack insert of
the Reflotron test.
Notes on result displays in special cases
Checking results