2.22
In the case of non-linear
kinetics and a value at the
lower end of the measuring
range, there is no purpose in
diluting the sample.
The reaction kinetics were
strongly non-linear.
Repeat the test.
However, it is possible that the
sample cannot be determined
with Reflotron.
The result remains displayed
until the flap is opened.
Note
It is possible to have a second
look at the results later on.
R see “Storage and retrieval of
results” in Section 2.
Remove the used test strip and
inspect the colour development
of the test area.
15 . 0 5 . 01 09 :13 : 0 4
IN V A L I D R E S U L T
15 . 0 5 . 01 09 :13 : 0 6
RE A C T I ON N O N L INE A R
15 . 0 5 . 01 09 :13 : 0 4
GL U 12 8 . 0 m g/ d l
Qq
Opening the measuring
chamber and removing the
strip