5.3
Operator messages
Meaning
The measuring chamber has
been opened. A used test strip
has been left in the measuring
chamber.
Reflotron Plus received no data
from the test strip after the
measuring chamber was
closed.
The data read from the test
strip are incomplete or faulty.
The measurement has been
interrupted because the meas-
uring chamber was opened too
soon.
The measurement has been
interrupted because the meas-
uring chamber was opened and
closed too soon.
More than 7 days have elapsed
or more than 100 measure-
ments have been taken since
the measuring chamber was
last cleaned.
Action
Remove the test strip. If you
find the chamber empty, R
push the release lever upwards.
Insert the test strip with the
brown magnetic strip facing
down and the test area pointing
forward.
Insert the test strip with the
brown magnetic strip facing
down and the test area pointing
forward. Important: Do not
store the test strips near mag-
netic objects (staple magnets,
magnetized paper clip holders,
etc).
Close the flap and wait for the
next message.
Open the flap, remove and
discard the test strip.
Clean R see Section 4
“Cleaning and Quality
Checking”.
Message
15 . 0 5 . 01 12 : 0 9 : 56
PL E A S E RE M O V E STR I P
15 . 0 5 . 01 12 : 0 9 : 56
NO M A G N ET I C C O DE !
15 . 0 5 . 01 12 : 0 9 : 56
MA G N E T IC C O D E MIS R E A D !
15 . 0 5 . 01 12 : 0 9 : 56
ME A S U R EM E N T I NTE R R U P TE D !
followed by
15 . 0 5 . 01 12 : 0 9 : 56
PL E A S E CL O S E F LAP
15 . 0 5 . 01 12 : 0 9 : 56
ME A S U R EM E N T I NTE R R U P TE D !
followed by
15 . 0 5 . 01 12 : 0 9 : 56
PL E A S E RE M O V E STR I P
15 . 0 5 . 01 12 : 0 9 : 56
CL E A N A ND C H E C K