5.2
15 . 0 5 . 01 1 2: 0 9 : 5 6
GL U 128 . 0 m g /d l
15 . 0 5 . 01 1 2: 0 9 : 5 6
GL U < 1 0 . 0 m g / d l
15 . 0 5 . 01 1 2: 0 9 : 5 6
GL U > 6 0 0 m g /d l
15 . 0 5 . 01 1 2: 0 9 : 5 6
GG T 31 . 0 U / l 2 5 ° C
15 . 0 5 . 01 12 : 0 9 : 56
GG T *2 0 0 U / l 2 5 ° C
15 . 0 5 . 01 12 : 0 9 : 56
NO N L I NE A R R E ACT I O N !
15 . 0 5 . 01 12 : 0 9 : 56
IN V A L I D R E S U L T !
15 . 0 5 . 01 12 : 0 9 : 56
NO N L I NE A R R E ACT I O N !
15 . 0 5 . 01 12 : 0 9 : 56
GG T *2 0 0 U / l 3 7 ° C
15 . 0 5 . 01 12 : 0 9 : 56
OU T O F T ES T R A N GE !
15 . 0 5 . 01 12 : 0 9 : 56
GG T D I L UT E S A M PLE !
NO P A R A ME T E R
NO E N T R Y !
Display of the result
(substrates).
The result is below the
measurement range.
The result is above the
measurement range.
Display of result for enzymes.
(The chosen reference tempera-
ture is also displayed).
The reaction kinetics has
slightly deviated from the linear
course.
The reaction kinetics has
considerably deviated from
the linear course.
It was not possible to check the
reaction kinetics for linearity.
Reason: very high enzyme activ-
ity/concentration or intense intrin-
sic colour of sample material.
The concentration/enzyme
activity is too high or the intrin-
sic colour of the sample mate-
rial too intense.
The unit or enzyme reference
temperature cannot be altered,
as no test strip has been mea-
sured.
Indicates that no further record
has been found when retrieving
results from memory, when
using the “search and copy”
function or in the transfer or
print profile functions (see
Section 3).
Record the result.
If the result appears implaus-
ible, check whether insufficient
or no sample material was
applied.
If possible, dilute the sample as
specified. R See pack insert of
Reflotron tests for details.
Record the result.
Repeat the test as directed
(possibly after diluting the
sample). R pack insert of
Reflotron tests.
Repeat the test as directed.
Determination of this sample
may not be possible.
Repeat the test as directed,
possibly after diluting the
sample.
Dilute sample as directed. R
See pack insert of Reflotron
tests.
Take a measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Reflotron Plus and is the answer not in the manual?

Roche Reflotron Plus Specifications

General IconGeneral
BrandRoche
ModelReflotron Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals