Checking the whole system with control sera
The following control sera are available for checking the Reflotron
Plus, the Reflotron tests and operation of the instrument:
Reflotron Precinorm U, the universal system control;
Reflotron Precinorm HB, a special control material for Reflotron
Hemoglobin;
Reflotron Precinorm HDL, a special control material for Cholesterol
HDL.
Control sera should be handled as potentially infectious materials.
Observe the relevant regulations and safety measures applicable to
your laboratory or practice.
Carefully read the pack insert accompanying the control material.
Open the bottle, carefully avoiding any loss of lyophilized material
and, in accordance with the pack insert, add exactly the amount of
distilled water specified using a suitable pipette.
Reconstituting the control
material