4.6
Close the bottle carefully and dissolve its contents completely by
occasionally swirling and inverting the bottle as described in the pack
insert, avoiding the formation of foam. Wait until the stated reconsti-
tution time has elapsed.
The solution is then used for the analysis as a sample.
For stability, see the pack insert.
Compare the result with the values specified in the table supplied
with the control material. If it lies outside the specified confidence
limits, proceed as follows:
Check the expiry date of the test strips and the lyophilized or recon-
stituted control material.
Check that you are handling the pipette (or applicator) and the test
correctly.
Clean the measuring chamber.
Check the performance of the optical system with Reflotron Check.
Repeat the test with the control serum.
Evaluation