4.1
4. Cleaning and Quality Checking
Cleaning the instrument
Note
Always switch off the instrument prior to cleaning.
Do not allow fluids to enter the instrument.
Some parts of the instrument can become contaminated with infec-
tious material. Please observe the pertinent safety regulations.
The Reflotron Plus needs no maintenance.
The exterior parts of the instrument can be cleansed as required
using mild detergents or disinfectants, preferably 70% alcohol.
After 100 measurements or 7 days after being cleaned last,
Reflotron Plus automatically asks the user to clean the measuring
system and subsequently to check the optical system R see also
“Display Messages and What They Mean”:
Improper (excessive) application of the sample can make it necessary
to carry out cleaning at considerably shorter intervals!
Allow Reflotron Plus to stand for at least 10 minutes after completion
of the cleansing operations before switching on again!
Do not spray disinfectants into the measuring chamber.
Open the flap to clean the measuring chamber.
If there is still a test strip in the measuring chamber this should be
removed. If it cannot be removed easily, lift up the release lever and
then remove the test strip.
Opening the measuring
chamber
15 . 0 5 . 01 12 :09 : 5 6
CL E A N A ND C H E C K

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Reflotron Plus and is the answer not in the manual?

Roche Reflotron Plus Specifications

General IconGeneral
BrandRoche
ModelReflotron Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals