Note
Keep the Reflotron test strip
away from magnetic fields
(e.g. paper clip holders,
magnets, magnetic stirrers) as
these can corrupt or delete the
magnetic code. This is, how-
ever, reliably recognized by the
instrument.
The magnetic code has not
been read.
Reason: the test strip has not
been inserted correctly.
In this case take a new test
strip and insert it with its mag-
netic strip facing down-
wards and its test area at the
front. If necessary, check the
transporter and the magnetic
head for soiling as described in
Section 4 “Cleaning and Quality
Checking”.
These messages appear if the
flap was opened during
measurement. This stops the
measurement, and no result is
therefore displayed.
If this happens, close the flap,
remove the test strip and
repeat the test with a new test
strip.
Repetition of measurement with
the same test strip returns
unusable results.
The instrument displays the
result along with the parameter
abbreviation and the units.
In enzyme determinations the
reference temperature is also
displayed.
15 . 0 5 . 01 09 :10 : 4 0
NO M A G N ET I C C O DE
15 . 0 5 . 01 09 :10 : 5 2
PL E A S E CL O S E F LAP
15 . 0 5 . 01 09 :13 : 0 4
GL U 1 2 8 .0 m g / d l
15 . 0 5 . 01 09 :13 : 0 4
GG T 31 . 0 U / l 3 7 ° C
15 . 0 5 . 01 09 :10 : 5 0
ME A S U R EM E N T I NTE R R U P TE D
Display of results