2.19
Close the flap.
The magnetic code is read.
The appearance of the parame-
ter abbreviation “GLUconfirms
that the Reflotron Glucose
test strip has been correctly
inserted and the data of its
magnetic code have been read
by the microprocessor.
The same abbreviation (in this
case GLU”) also appears on
the label of the test strip con-
tainer and on the “handleof
each Reflotron test strip for
purposes of identification.
The time remaining before the
result appears is displayed in
seconds.
If the instrument displays one of
the following messages instead
of GLU RESULT IN 143 SEC”,
then something is wrong:
The magnetic code was not
completely read. Remove the
test strip as soon as the
message appears, insert it
again correctly, and close the
flap once again. If the same
message appears, repeat the
test with a new test strip.
Closing the flap
Other messages
15 . 0 5 . 01 09 :10 : 4 2
RE A D I N G M A G N E TIC C O D E
15 . 0 5 . 01 09 :10 : 4 7
GL U R E S UL T I N 1 43 s e c
15 . 0 5 . 01 09 :10 : 4 0
MA G N E T IC C O D E MIS R E A D