6.16
Over 50 function controls ensure the working safety of the instru-
ment.
alphanumeric display; 2 lines with 24 characters each; LC display
battery-buffered real-time clock
± 0.5% of the reflectance with respect to the mean for the instru-
ments delivered.
0.2% of the reflectance
0.05% deviation
Temperature: +15 °C to +34 °C
Rel. humidity: max. 95%
Temperature: –20 °C to +55 °C when packed
Rel. humidity: 5% to 95% when packed
1 RS 232 C serial interface, DB-25 socket
1 keyboard interface, 5-pin DlN socket
115 230 V AC
(+/– 22%)
Frequency: 47 Hz to 63 Hz
Permissible range: 10–30 V DC
(reverse battery protection)
Permissible internal resistance: max. 100 m)
Recommended plug: Cannon AXR-3-11
Connecting cable: max. 2.5 m; 1.5 mm
45 VA
Display:
Clock:
6. Accuracy
7. Precision
8. Linearity
9. Required operating
environment
10. Transport conditions
11. Data interfaces
12. Power supply
Mains:
Car battery (optional):
13. Power consumption

Table of Contents