2.13
Reflotron Pipette Correct application of the
sample material
When working with Reflotron
test strips a specimen volume
of 30 µl must be applied with a
commercially available precision
pipette. The Reflotron Pipette is
also suitable for this purpose.
Reflotron Pipette has been
developed for general use with
undiluted blood, serum and
plasma. It delivers precisely
28.5 – 31.5 µl, the volume
needed for correct functioning
of the Reflotron Plus system.
In the case of capillary blood,
certain precautions with respect
to the specimen collection must
be observed (see so “Working
with capillary blood”).