2.27
Applying the sample
Interim storage of the
specimens
If several determinations are to be carried out on the same patient,
several capillary pipettes containing capillary blood can be stored for
a few minutes prior to commencing the respective measurements.
The capillary pipettes contained in the Accessory Kit are coated with
lithium heparinate. Please note that not every test can be performed
with this anticoagulant. R See the pack insert of the respective
Reflotron test.
The dispensing button of the applicator has two stops.
By pressing the button until the first stop is reached, the sample is
applied to the centre of the mesh-covered application zone (see also
illustration on page 2.17). Hold the applicator vertically while dis-
pensing and perform the action slowly in order to avoid splattering.
The volume of the capillary pipettes is such that sufficient sample is
applied to the test strip.
After the sample has been dispensed, the capillary tube can be eject-
ed by depressing the dispensing button until the second stop is
reached.
Dispose of potentially infectious material according to the regulations
applicable in your practice or laboratory.
Discarding the capillary
pipette

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche Reflotron Plus and is the answer not in the manual?

Roche Reflotron Plus Specifications

General IconGeneral
BrandRoche
ModelReflotron Plus
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals