2.37
Changing the substrate unit
UN I T S T RI G m g/ d l m m ol/ l
TE M P G P T 2 5 ° C 3 0°C 3 7 ° C
UN I T S G PT U / l
µ
ka t / l
Changing the enzyme
reference temperature
If the last test performed was
an enzyme determination, e.g.
GPT, the desired temperature
can be selected using the <”
and “>” keys. After confirmation
with the prompt for the
desired unit appears.
µkat/l stands for
microkatals/litre, U/l for units
per litre.
Use the “<” or “>key to posi-
tion the cursor under the
desired unit and confirm
with .
If the last test performed was
a substrate measurement, e.g.
“TRIGLYCERIDE”, the desired
unit can be selected with the
keys <” and “>”. Confirm by
pressing ”.
Press the two buttons “<” and
“>simultaneously to return
to measuring mode. The
Reflotron Plus is now ready to
perform an analysis again. See
also R Basic Instrument
Settings in Section 6.