4.3
Hold the black shield forwards and pull upwards, pulling the bridge
upwards at the same time.
Also wipe the transporter, especially the lower heater and the mag-
netic head, e.g. with an alcohol-moistened wipe or a moist cotton
bud. In the latter case dry with a dry cotton bud.
If cotton buds are used for cleaning, contaminated parts can be dis-
infected by subsequent cleaning with a wet wipe (70 % alcohol), e.g.
from Reflotron Clean and Check.
Then check the correct position of the lower heater: springing, free
moving, parallel to the transporter surface.
If this is not the case, contact the service department of Roche
Diagnostics.
Return the black shield to its original position.
Leave the instrument to dry for at least 10 minutes with the measur-
ing chamber flap open.
Close the flap.
After every cleaning operation, the optical system of Reflotron
Plus should be checked using Reflotron Check (REF 11142577).