Remote Control Commands
R&S
®
NRPxxA(N)
103User Manual 1177.6017.02 ─ 05
R&S NRP2 base unit
S-Parameter Device Locked ← Global Flags
If enabled, the S-parameter device that is selected as the default device in the table of
S-parameter devices is locked and cannot be changed using:
[SENSe<Sensor>:]CORRection:SPDevice:SELect
R&S NRP2 base unit
The default S-parameter device is the S-parameter device that you have selected
when the power sensor is started.
Use Flags in Factory Cal. Data Set ← Global Flags
Available if the power sensor supports two different calibration data sets:
Factory calibration data set containing all factory calibration data.
User calibration data set that contains the S-parameter devices you have loaded.
Note: After you have added S-parameter devices and configured the global flags, dis-
able this option. Otherwise, it is not possible to enable S-parameter correction because
the flags in the factory calibration data set prevent it.
Device table
Shows a list of all S-parameter devices that are available in the calibration data set.
If you double-click an entry, a dialog for the device is opened that allows to import,
export, and edit S-parameter data. See "To change the S-parameter data"
on page 105.