Querying the register:
STATus:QUEStionable:CONDition?
STATus:QUEStionable[:EVENt]?
Table 11-6: Meaning of bits used in the questionable status register
Bit No. Meaning
0 to 2 Not used
3 Questionable power status register summary bit
Corresponds to the summary bit of the questionable power status register.
See Chapter 11.3.6.1, "Questionable Power Status Register", on page 150.
4 to 7 Not used
8 Questionable calibration status register summary bit
Corresponds to the summary bit of the questionable calibration status register.
See Chapter 11.3.6.2, "Questionable Calibration Status Register", on page 150.
9 POST failure
The built-in test of the R&S NRPxxA(N) carried out automatically upon power-up has
generated an error.
10 to 15 Not used
11.3.6.1 Questionable Power Status Register
The CONDition register contains information whether the measured power values are
questionable.
Querying the register:
STATus:QUEStionable:POWer:CONDition?
STATus:QUEStionable:POWer[:SUMMary][:EVENt]?
Table 11-7: Meaning of bits used in the questionable power status register
Bit No. Meaning
0 Not used
1 Sensor power
The measurement data of the sensor is corrupt.
2 to 4 Not used
5 Sensor please zero
The zero correction for the sensor is no longer correct and should be repeated.
6 to 15 Not used
11.3.6.2 Questionable Calibration Status Register
The EVENt register and the CONDition register contain information whether the zero
offset of a sensor is still valid.