Remote Control Basics
R&S
®
NRPxxA(N)
153User Manual 1177.6017.02 ─ 05
Table 11-10: Meaning of bits used in the operation status register
Bit No. Meaning
0 Calibrating status register summary bit
This bit is set if the sensor is being calibrated.
See Chapter 11.3.8.1, "Operation Calibrating Status Register", on page 153.
1 to 3 Not used
4 Measuring status register summary bit
This bit is set if the sensor is performing a measurement.
See Chapter 11.3.8.2, "Operation Measuring Status Register", on page 154.
5 Trigger status register summary bit
This bit is set if the sensor is in the waiting for trigger state, i.e. waiting for a trigger event.
See Chapter 11.3.8.3, "Operation Trigger Status Register", on page 154.
6 Not used
7 to 9 Not used
10 Sense status register summary bit
This bit is set if a sensor is initialized.
See Chapter 11.3.8.4, "Operation Sense Status Register", on page 155.
11 Lower limit fail status register
This bit is set if a displayed value has dropped below a lower limit value.
See Chapter 11.3.8.5, "Operation Lower Limit Fail Status Register", on page 155.
12 Upper limit fail status register
This bit is set if a displayed value has exceeded an upper limit value.
See Chapter 11.3.8.6, "Operation Upper Limit Fail Status Register", on page 156.
13 to 14 Not used
15 Bit 15 is never used.
11.3.8.1 Operation Calibrating Status Register
The CONDition register contains information about whether a sensor is calibrated
and, depending on the configuration of the transition register. The EVENt register indi-
cates whether a calibration was started or completed since the last readout of this reg-
ister.
Querying the register:
●
STATus:OPERation:CALibrating:CONDition?
●
STATus:OPERation:CALibrating[:SUMMary][:EVENt]?
Status Reporting System