EasyManua.ls Logo

Rohde & Schwarz NRPxxA Series - Configuring and Retrieving Results; Setting the Power Unit; Trigger:sync:port; Trigger:sync:state

Rohde & Schwarz NRPxxA Series
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
NRPxxA(N)
83User Manual 1177.6017.02 ─ 05
Parameters:
<source> HOLD | IMMediate | INTernal | BUS | EXTernal | EXT1 |
EXTernal1 | EXT2 | EXTernal2
See Chapter 9.5.2.3, "Trigger Sources", on page 73.
*RST: IMMediate
Manual operation: See "<Source>" on page 52
TRIGger:SYNC:PORT <sync_port>
Selects the external connection for the sync output of the sensor. For more information,
see TRIGger:SYNC:STATe.
Parameters:
<sync_port> EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST: EXT1
TRIGger:SYNC:STATe <state>
Usually used in combination with TRIGger:MASTer:STATe ON.
If enabled, blocks the external trigger bus as long as the sensor remains in the mea-
surement state. Thus, ensures that a new measurement is only started after all sen-
sors have completed their measurements.
Make sure that the number of repetitions is the same for all sensors involved in the
measurement. Otherwise, the trigger bus is blocked by any sensor that has completed
its measurements before the others and has returned to the idle state.
Parameters:
<state> *RST: OFF
9.6 Configuring and Retrieving Results
Setting the Power Unit............................................................................................ 83
Setting the Result Format....................................................................................... 84
Retrieving Results...................................................................................................85
9.6.1 Setting the Power Unit
The UNIT subsystem contains commands for setting up the power unit.
UNIT:POWer <unit>
Sets the output unit for the measured power values.
Configuring and Retrieving Results

Table of Contents

Related product manuals