Remote Control Basics
R&S
®
NRPxxA(N)
155User Manual 1177.6017.02 ─ 05
11.3.8.4 Operation Sense Status Register
The CONDition register contains information about whether a sensor is being initial-
ized and, depending on the configuration of the transition register. The EVENt register
indicates whether a sensor initialization was started or completed since the last read-
out of this register.
This status is assumed by a sensor if one of the following conditions is met:
●
The supply voltage is switched on (power up)
●
The sensor was connected
●
A reset was performed using:
– *RST
– SYSTem:PRESet
Querying the register:
●
STATus:OPERation:SENSe:CONDition?
●
STATus:OPERation:SENSe[:SUMMary][:EVENt]?
Table 11-14: Meaning of bits used in the operation sense status register
Bit No. Meaning
0 Not used
1 Sensor initializing
The sensor is being initialized.
5 to 15 Not used
11.3.8.5 Operation Lower Limit Fail Status Register
The CONDition register contains information about whether a displayed value is
below a configured lower limit. The EVENt register indicates whether a measured
value dropped below a limit value since the last readout of the Operation Lower Limit
Fail Status Register. Details of the behavior are defined by the transition register.
Querying the register:
●
STATus:OPERation:LLFail:CONDition?
●
STATus:OPERation:LLFail[:SUMMary][:EVENt]?
Table 11-15: Meaning of bits used in the operation lower limit fail status register
Bit No. Meaning
0 Not used
1 Lower limit fail
The measured value drops below the lower limit value.
5 to 15 Not used
Status Reporting System