EasyManua.ls Logo

Rohde & Schwarz NRPxxA Series - Operation Upper Limit Fail Status Register

Rohde & Schwarz NRPxxA Series
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Basics
R&S
®
NRPxxA(N)
156User Manual 1177.6017.02 ─ 05
11.3.8.6 Operation Upper Limit Fail Status Register
The CONDition register contains information about whether a displayed value is
above a configured upper limit. The EVENt register indicates whether a limit value was
exceeded since the last readout of the Operation Upper Limit Fail Status Register.
Querying the register:
STATus:OPERation:ULFail:CONDition?
STATus:OPERation:ULFail[:SUMMary][:EVENt]?
Table 11-16: Meaning of bits used in the operation lower limit fail status register
Bit No. Meaning
0 Not used
1 Upper limit fail
The measured value exceeds the upper limit value.
5 to 15 Not used
Status Reporting System