6.5.2 Averaging Settings
Describes the parameters for automatic averaging.
Further information:
●
Chapter 9.8.1, "Configuring Auto Averaging", on page 90
Access: main dialog of the web user interface > navigation pane > "Averaging"
Averaging Mode............................................................................................................50
└ <Mode>...........................................................................................................50
└ <Value>...........................................................................................................50
Filter Terminal Control...................................................................................................51
Auto Measurement Time...............................................................................................51
Averaging Mode
Groups the averaging settings. See also Chapter 9.8.1, "Configuring Auto Averaging",
on page 90.
<Mode> ← Averaging Mode
Sets the averaging mode.
"manual"
Disables auto averaging. Enter the average count under "<Value>"
on page 50.
"auto 1 dB" / "auto 0.1 dB" / "auto 0.01 dB" / "auto 0.001 dB"
Uses an automatic averaging filter with the respective resolution
index.
"noise content"
Predefines the compliance to an exactly defined noise component.
Enter this value under "<Value>" on page 50.
Remote command:
[SENSe<Sensor>:]AVERage[:STATe] on page 94
[SENSe<Sensor>:]AVERage:COUNt:AUTO on page 91
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE on page 93
<Value> ← Averaging Mode
The content of this field depends on the setting under "<Mode>" on page 50.
●
If "manual" is set:
Sets the average count, also called averaging factor.
●
If "auto xx dB" is set:
Displays the resolution index.
●
If "noise content" is set:
Settings