Remote Control Commands
R&S
®
NRPxxA(N)
123User Manual 1177.6017.02 ─ 05
Example:
SYST:SERR:LIST?
Query
0,"reported at uptime:2942; notice;
auto-averaging exceeded maximum time;
Notification"
Response
Usage: Query only
SYSTem:SUTime <update_time>
Effective only in the NRP legacy mode.
Sets the status update time. That is the maximum rate in which the power sensor can
output measurement results. Relevant only in continuous measurement mode
(INITiate:CONTinuousON).
Parameters:
<update_time> Range: 0.0 to 10.0
*RST: 10e-3
Default unit: Seconds
SYSTem:TLEVels?
Queries the possible power test levels of the sensor.
Usage:
Query only
SYSTem:TRANsaction:BEGin
Starts a series of settings.
Usage:
Event
SYSTem:TRANsaction:END
Ends a series of settings.
Usage:
Event
SYSTem[:SENSor]:NAME <sensorname>
Sets the name of the sensor according to your requirements. The specified name is
displayed in the web user interface of the network sensors.
The sensor name that you specify here is independent from the hostname of the sen-
sor. However, if the sensor name is not specified, it defaults to the hostname.
Example:
SYST:NAME "InputModule-X5"
Configuring the System