Remote command:
[SENSe<Sensor>:][POWer:][AVG:]APERture on page 88
Duty Cycle
Sets the duty cycle, the percentage of one period during which the signal is active, for
pulse modulated signals. If the duty cycle is set, the sensor calculates the signal pulse
power from its value and the average power.
Remote command:
[SENSe<Sensor>:]CORRection:DCYCle:STATe on page 95
[SENSe<Sensor>:]CORRection:DCYCle on page 95
Smoothing
Enables the smoothing filter, a steep-cut off digital lowpass filter. The filter reduces
result fluctuations caused by modulation.
Remote command:
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe on page 90
6.5 Settings
Describes the parameters for general sensor configuration.
Sensor Settings.......................................................................................................47
Averaging Settings..................................................................................................50
Trigger Settings.......................................................................................................51
System Settings...................................................................................................... 53
6.5.1 Sensor Settings
Describes the parameters for optimizing the measurement results for specific measure-
ment requirements.
Further information:
Chapter 9.4.2, "Selecting a Measurement Path", on page 69
Chapter 9.8.3, "Configuring Corrections", on page 94
Chapter 9.9, "Calibrating/Zeroing the Power Sensor ", on page 109
Chapter 9.10, "Testing the Power Sensor", on page 111
Access: main dialog of the web user interface > navigation pane > "Sensor"