Browser-Based User Interface
R&S
®
NRPxxA(N)
48User Manual 1177.6017.02 ─ 05
Range............................................................................................................................48
Γ Correction...................................................................................................................48
└ <State>........................................................................................................... 48
└ Magnitude....................................................................................................... 48
└ Phase..............................................................................................................48
Zero Calibration.............................................................................................................49
Diagnostics....................................................................................................................49
Range
Selects which path of the sensor is used for the measurement.
Remote command:
[SENSe<Sensor>:]RANGe:AUTO on page 70
[SENSe<Sensor>:]RANGe on page 70
Γ Correction
Groups the parameters for the complex reflection coefficient. See also Chapter 9.8.3.4,
"S-Gamma Corrections", on page 98.
<State> ← Γ Correction
Enables or disables the use of the complex reflection coefficient of the signal source,
Γ
source
.
Remote command:
[SENSe<Sensor>:]SGAMma:CORRection:STATe on page 99
Magnitude ← Γ Correction
Sets the magnitude of the complex reflection coefficient of the source , Γ
source
.
A value of 0.0 corresponds to an ideal matched source. A value of 1.0 corresponds to
total reflection.
Remote command:
[SENSe<Sensor>:]SGAMma:MAGNitude on page 99
Phase ← Γ Correction
Sets the phase angle of the complex reflection coefficient of the source, Γ
source
.
Remote command:
[SENSe<Sensor>:]SGAMma:PHASe on page 99