Remote Control Commands
R&S
®
NRPxxA(N)
99User Manual 1177.6017.02 ─ 05
Sensor
Source
S
21
S
12
S
22
S
11
NRP
3-Path Diode Power Sensor
MHz to GHz, 100 pW to 200 mW (70 dBm to +23 dBm)
SMART SENSOR TECHNOLOGY
Figure 9-4: Correction of interactions between the power sensor, the signal source, and the S-param-
eter device
Remote commands:
[SENSe<Sensor>:]SGAMma:CORRection:STATe................................................................99
[SENSe<Sensor>:]SGAMma:MAGNitude........................................................................... 99
[SENSe<Sensor>:]SGAMma:PHASe................................................................................. 99
[SENSe<Sensor>:]SGAMma:CORRection:STATe <state>
Enables or disables the use of the complex reflection coefficient to correct the interac-
tions of power sensor and signal source.
Parameters:
<state> *RST: OFF
Manual operation: See "<State>" on page 48
[SENSe<Sensor>:]SGAMma:MAGNitude <magnitude>
Sets the magnitude of the complex reflection coefficient of the source, Γ
source
.
Parameters:
<magnitude> 0.0 ideal matched source
1.0 total reflection
Range: 0.0 to 1.0
*RST: 0.0
Manual operation: See "Magnitude" on page 48
[SENSe<Sensor>:]SGAMma:PHASe <phase>
Sets the phase angle of the complex reflection coefficient of the source, Γ
source
.
Parameters:
<phase> Range: -360.0 to 360.0
*RST: 0.0
Manual operation: See "Phase" on page 48