EasyManua.ls Logo

Rohde & Schwarz NRPxxA Series - Trigger:master:state; Trigger:slope; Trigger:source

Rohde & Schwarz NRPxxA Series
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
NRPxxA(N)
82User Manual 1177.6017.02 ─ 05
If the sensor triggers itself, the trigger source of the sensor must be assigned to the
other external port, as shown in the examples.
Parameters:
<master_port> EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST: EXT1
Example:
TRIG:MAST:PORT EXT1
TRIG:SOUR EXT2
TRIG:MAST:STAT ON
Example:
TRIG:MAST:PORT EXT2
TRIG:SOUR EXT1
TRIG:MAST:STAT ON
TRIGger:MASTer:STATe
<state>
Enables or disables the trigger master mode of the sensor. In this state, the power sen-
sor can output a digital trigger signal in sync with its own trigger event.
If enabled, select the output port for the trigger signal using TRIGger:MASTer:PORT.
Typically, the trigger master uses its internal trigger source. But you can also trigger the
trigger master externally, because the power sensor has got two external trigger con-
nectors. If you trigger the trigger master externally, use EXTernal1 as external trigger
input port (trigger source) and EXTernal2 as trigger master output port or vice versa.
Parameters:
<state> *RST: OFF
TRIGger:SLOPe <slope>
Effective only if TRIGger:SOURce is set to INTernal or EXTernal.
Determines which edge of the envelope power, with internal triggering, or increasing
voltage, with external triggering, is used for triggering.
Parameters:
<slope> POSitive | NEGative
POSitive
Rising edge
NEGative
Falling edge
*RST: POSitive
Manual operation: See "<Slope>" on page 52
TRIGger:SOURce <source>
Selects the source for the trigger event detector.
Controlling the Measurement

Table of Contents