[SENSe<Sensor>:][POWer:][AVG:]FAST <state>
Enables or disables a fast unchopped Continuous Average measurement. The aver-
age count is enforced to 1. Any setting for average count is silently ignored. See also
"Calculating the measurement time" on page 88.
Parameters:
<state> *RST: OFF
Example: See Chapter 10, "Performing Measurement Tasks - Program-
ming Examples", on page 128.
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe <state>
Enables or disables the smoothing filter, a steep-edge digital lowpass filter. If you can-
not adjust the aperture time exactly to the modulation period, the filter reduces result
fluctuations caused by modulation.
Parameters:
<state> ON | OFF
*RST: OFF
Example:
SMO:STAT OFF
Manual operation: See "Smoothing" on page 47
9.8 Configuring Basic Measurement Parameters
The following section describes the settings common for several measurement modes.
9.8.1 Configuring Auto Averaging
Describes the commands for automatic averaging in the continuous average measure-
ments.
Remote commands:
[SENSe<Sensor>:]AVERage:COUNt..................................................................................91
[SENSe<Sensor>:]AVERage:COUNt:AUTO........................................................................91
[SENSe<Sensor>:]AVERage:COUNt:AUTO:MTIMe.............................................................91
[SENSe<Sensor>:]AVERage:COUNt:AUTO:NSRatio...........................................................92
[SENSe<Sensor>:]AVERage:COUNt:AUTO:RESolution.......................................................92
[SENSe<Sensor>:]AVERage:RESet...................................................................................92
[SENSe<Sensor>:]AVERage:COUNt:AUTO:TYPE.............................................................. 93
[SENSe<Sensor>:]AVERage:TCONtrol.............................................................................. 93
[SENSe<Sensor>:]AVERage[:STATe].................................................................................94
Configuring Basic Measurement Parameters