EasyManua.ls Logo

Rohde & Schwarz NRPxxA Series - Operation Measuring Status Register; Operation Trigger Status Register

Rohde & Schwarz NRPxxA Series
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Basics
R&S
®
NRPxxA(N)
154User Manual 1177.6017.02 ─ 05
Table 11-11: Meaning of bits used in the operation calibrating status register
Bit No. Meaning
0 Not used
1 Sensor calibrating
The sensor is being calibrated
5 to 15 Not used
11.3.8.2 Operation Measuring Status Register
The CONDition register contains information about whether a measurement is being
performed by a sensor and, depending on the configuration of the transition register.
The EVENt register indicates whether a measurement was started or completed since
the last readout of this register.
Querying the register:
STATus:OPERation:MEASuring:CONDition?
STATus:OPERation:MEASuring[:SUMMary][:EVENt]?
Table 11-12: Meaning of bits used in the operation measuring status register
Bit No. Meaning
0 Not used
1 Sensor measuring
The sensor is performing a measurement.
5 to 15 Not used
11.3.8.3 Operation Trigger Status Register
The CONDition register contains information about whether a sensor is in the waiting
for trigger state, i.e. expecting a trigger event and, depending on the configuration of
the transition register. TheEVENt register indicates whether the waiting for trigger state
was entered or quit by a sensor since the last readout of the register.
Querying the register:
STATus:OPERation:TRIGger:CONDition?
STATus:OPERation:TRIGger[:SUMMary][:EVENt]?
Table 11-13: Meaning of bits used in the operation trigger status register
Bit No. Meaning
0 Not used
1 Sensor waiting for trigger
The sensor is in the waiting for trigger state and is waiting for a trigger event. When the
trigger event occurs, the sensor changes into the measuring state.
5 to 15 Not used
Status Reporting System

Table of Contents

Related product manuals