152User Manual 1177.6017.02 ─ 05
Bit No. Meaning
4 Execution error
This bit is set if the syntax of a received command is correct but the command cannot be
executed due to various marginal conditions. An error message with a number between
-200 and -300 denoting the error in greater detail is entered in the error queue.
5 Command error
This bit is set if an undefined command or a command with incorrect syntax is received. An
error message with a number between -100 and -200 denoting the error in greater detail is
entered in the error queue.
6 User request
This bit is set when the sensor is switched over to manual control.
7 Power on
This bit is set when the sensor is switched on.
11.3.8 Operation Status Register
The CONDition register contains information on the operations currently being per-
formed by the sensor, while the EVENt register contains information on the operations
performed by the sensor since the last readout of the register.
1
0
2
3
4
5
6
7
9
8
10
11
12
13
14
15
+
Operation
Status
Calibrating
0
0
0
Measuring
Triggering
0
0
0
0
Sense Summary
Lower Limit Fail
Upper Limit Fail
0
0
0
Querying the register:
STATus:OPERation:CONDition?
STATus:OPERation[:EVENt]?