Browser-Based User Interface
R&S
®
NRPxxA(N)
49User Manual 1177.6017.02 ─ 05
Zero Calibration
Performs zeroing using the signal at the sensor input. See Chapter 9.9, "Calibrating/
Zeroing the Power Sensor ", on page 109.
Note:
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
Remote command:
CALibration<Channel>:ZERO:AUTO on page 110
Diagnostics
Triggers a selftest of the sensor.
Note:
Do not apply a signal to the sensor while the selftest is running. If the selftest is carried
out with a signal being present, error messages can erroneously be output for the test
steps Offset Voltages and/or Noise Voltages.
The results of the selftest are shown in a dialog that is displayed after the test comple-
tion.
Remote command:
TEST:SENSor? on page 111