EasyManua.ls Logo

Rohde & Schwarz NRPxxA Series - Page 60

Rohde & Schwarz NRPxxA Series
165 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Firmware Update
R&S
®
NRPxxA(N)
60User Manual 1177.6017.02 ─ 05
Example:
You want to update your R&S NRP18AN with the nrp6an_FW_15.02.12.01.rsu
file. This file has a size of 10242884 bytes.
To send the file to the sensor for updating the firmware, your application has to assem-
ble a memory block containing:
SYST:FWUP <block_data>
The <block_data> are definite length arbitrary block data as described in SYSTem:
FWUPdate on page 117.
The size of the file is 10242884. This number has 8 digits. Thus, the <block_data>
consist of the following:
#
8
How many digits follow to specify the file size.
10242884
Number that specifies the file size.
<file_contents>
Contents of the *.rsu file, byte-by-byte
0x0a
Delimiter
In this example, you write exactly 10242905 bytes to the power sensor, for example by
using a 'viWrite()' function.
The 10242905 bytes result from the values of the list above:
9 + 1 + 1 + 1 + 8 + 10242884 + 1
In a (pseudo) string notation, the memory block looks as follows:
SYST:FWUP #810242884<file_contents>0x0a,
Updating the Firmware

Table of Contents

Related product manuals