EasyManua.ls Logo

Rohde & Schwarz SMA100B - Source > External Settings

Rohde & Schwarz SMA100B
778 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Analog Modulations
R&S
®
SMA100B
102User Manual 1178.3834.02 ─ 05
4.4.6.2 Source > External Settings
Access:
Select "Modulation" > "Modulation Sources > External".
The "External" settings section contains all parameters required to configure the
signal of an externally supplied LF signal.
Settings
Coupling (AC/DC)....................................................................................................... 102
Impedance.................................................................................................................. 102
Bandwidth................................................................................................................... 103
Show Connector..........................................................................................................103
Coupling (AC/DC)
Selects the coupling mode (AC or DC) for the external signal.
"AC"
Disconnects the DC voltage component and uses only the AC com-
ponent of the modulation signal.
"DC"
Uses the modulation signal with both components, AC and DC.
For active external exponential AM, sets the coupling of the corre-
sponding input automatically to DC.
Remote command:
[:SOURce<hw>]:INPut:MODext:COUPling<ch> on page 598
Impedance
Sets the impedance for the externally supplied signal.
Remote command:
[:SOURce<hw>]:INPut:MODext:IMPedance<ch> on page 599
Modulation Settings

Table of Contents

Other manuals for Rohde & Schwarz SMA100B

Related product manuals