Remote Control Commands
R&S
®
SMA100B
551User Manual 1178.3834.02 ─ 05
[:SOURce<hw>]:ADF:COMid:SYMBol <Symbol>
Sets the length of the Morse code symbol space.
Parameters:
<Symbol> float
Range: 50E-3 to 1
Increment: 1E-4
*RST: 100E-3
Example: See Example "Configure and generate a ADF signal"
on page 542.
Manual operation: See " Symbol Space " on page 152
[:SOURce<hw>]:ADF:COMid:TSCHema
<TSCHema>
Sets the time schema of the Morse code for the COM/ID signal.
Parameters:
<TSCHema> STD | USER
*RST: STD
Example: See Example "Configure and generate a ADF signal"
on page 542.
Manual operation: See "Time Schema" on page 152
[:SOURce<hw>]:ADF:COMid[:STATe] <State>
Enables/disables the COM/ID signal.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example: See Example "Configure and generate a ADF signal"
on page 542.
Manual operation: See "COM/ID State" on page 151
13.16.2.4 SOURce:ILS Subsystem
Option: R&S SMAB-K25
The SOURce:BB:ILS subsystem contains all commands for configuring the ILS signal.
Characteristics which are valid for all modulations and the LF output are configured in
the SOURce:LFOutput subsystem. The signal is output at the LF connector.
SOURce Subsystem