Improving Level Performance
R&S
®
SMA100B
236User Manual 1178.3834.02 ─ 05
"Float (Coarse/Fine)"
Auto scale is activated, that means this parameter automatically
selects the appropriate scaling of the y-axis so that the trace is
always visible. The range is either expanded, when a value is out of
the right/left end-of-scale value or it is reduced when the trace fits into
a smaller scale area. The step width is 5 dB for selection "Flt.
(Coarse)" and variable in the range of 0.2 db to 5 dB for selection "Flt.
(Fine)".
Remote command:
:SENSe[:POWer]:SWEep:FREQuency:YSCale:AUTO on page 468
:SENSe[:POWer]:SWEep:POWer:YSCale:AUTO on page 481
:SENSe[:POWer]:SWEep:TIME:YSCale:AUTO on page 486
Reset Auto Scale - Power Analysis
Resets the scaling of the y-axis to suitable values after the use of auto scaling in the
expanding mode. For this mode, the Y scale can get too expanded because of tempo-
rary high-power values. The reset function resets the diagram again, to indicate
smaller power values.
Remote command:
:SENSe[:POWer]:SWEep:FREQuency:YSCale:AUTO:RESet on page 469
:SENSe[:POWer]:SWEep:POWer:YSCale:AUTO:RESet on page 482
:SENSe[:POWer]:SWEep:TIME:YSCale:AUTO:RESet on page 487
Show Grid - Power Analysis
Activates the indication of a grid in the diagram area.
Remote command:
:DISPlay[:WINDow][:POWer]:SWEep:GRID:STATe on page 454
Background Color - Power Analysis
Sets the background color of the diagram to black or white. The background color is
also effective for the hardcopy of the diagram.
Remote command:
:DISPlay[:WINDow][:POWer]:SWEep:BACKground:COLor on page 453
Configure Gate Mode
For "Time" measurement mode only.
Access:
1. Select "Clk Syn/Power Sens" > "NRP Power Analysis".
2. Select "Configure" > "Configure Gate Mode"
Using Power Sensors