5. Set "Level > RF ON" to enable signal output.
4.6 How to Generate a Pulse Modulated Signal
The following example uses the internal pulse generator.
To set the frequency and level of the RF signal
1. Press [Preset].
2. In the status bar, set "Frequency = 4 GHz".
3. Set "Level = -25 dBm".
To configure the pulse generator
1. Select "Modulation" > "Pulse Generator".
2. Select "Pulse Mode = Double".
3. Set "Pulse Period = 10 us".
4. Set "Pulse Width = 2 us".
5. Set "Double Pulse Width = 1.2 us".
6. Set "Double Pulse Delay = 4.5 us".
7. Activate "Pulse Output State".
8. Set "Trigger Mode = Auto".
In the "Pulse Graph" tab, you can view the generated pulse signal graphically.
To enable pulse modulation
1. Select the "Pulse Modulation" tab.
2. Set "State = I" to activate pulse modulation.
3. Set "Level > RF ON" to enable signal output.
4.7 How to Generate a Pulse Train Modulated Signal
The following example uses the internal pulse generator.
To set the frequency and level of the RF signal
1. Press [Preset].
2. In the status bar, set "Frequency = 6 GHz".
3. Set "Level = -25 dBm".
How to Generate a Pulse Train Modulated Signal

Table of Contents

Other manuals for Rohde & Schwarz SMA100B

Related product manuals