Analog Modulations
R&S
®
SMA100B
89User Manual 1178.3834.02 ─ 05
Double Pulse Width
Sets the width of the second pulse.
Remote command:
[:SOURce<hw>]:PULM:DOUBle:WIDTh on page 532
Pulse Delay
Sets the pulse delay. The pulse delay determines the time that elapses after a trigger
event before pulse modulation starts. The pulse delay is not effective for double pulse
generation.
Remote command:
[:SOURce<hw>]:PULM:DELay on page 531
Double Pulse Delay
Sets the delay from the start of the first pulse to the start of the second pulse.
Remote command:
[:SOURce<hw>]:PULM:DOUBle:DELay on page 532
Pulse Output State
Activates the output of the pulse modulation signal.
Remote command:
[:SOURce<hw>]:PGENerator:OUTPut[:STATe] on page 627
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
Pulse Output Polarity
Sets the polarity of the pulse output signal.
Remote command:
[:SOURce<hw>]:PGENerator:OUTPut:POLarity on page 627
Execute Single Trigger
If "Trigger Mode = Single", initiates a single pulse sequence manually.
Remote command:
[:SOURce]:PULM[:INTernal][:TRAin]:TRIGger:IMMediate on page 531
4.4.3.2 Pulse Generator > Pulse Train Settings
Option: see Chapter 4.1, "Required Options", on page 73.
Access:
1. Select "Modulation" > "Pulse Modulation > Pulse Generator" > "Pulse Mode =
Train".