Remote Control Commands
R&S
®
SMA100B
534User Manual 1178.3834.02 ─ 05
[:SOURce<hw>]:PULM:POLarity <Polarity>
Sets the polarity of the externally applied modulation signal.
Parameters:
<Polarity> NORMal | INVerted
NORMal
Suppresses the RF signal during the pulse pause.
INVerted
Suppresses the RF signal during the pulse.
*RST: NORMal
Example: See Example "Perform pulse modulation" on page 529.
Manual operation: See "Polarity" on page 97
[:SOURce<hw>]:PULM:IMPedance <Impedance>
Sets the impedance for the external pulse trigger and pulse modulation input.
Parameters:
<Impedance> G50 | G10K
*RST: G50
Example: See Example "Perform pulse modulation" on page 529.
Manual operation: See "Impedance" on page 97
[:SOURce<hw>]:PULM:THReshold <Threshold>
Sets the threshold for the input signal at the [Pulse Ext] connector.
Parameters:
<Threshold> float
Range: 0 to 2
Increment: 0.1
*RST: 1
Default unit: V
Example:
SOURce:PULM:THReshold 1
Manual operation: See "Threshold" on page 97
Pulse Train Settings
Option: R&S SMAB-K27
This section describes the commands for the pulse train modulation and the associ-
ated file handling. For background information, see Chapter 4.4.3.2, "Pulse Generator
> Pulse Train Settings", on page 89.