8-13
Hazardous Area Oxymitter 5000
Figure 8-9. Fault 7, High Case Temp
Fault 7, High Case Temp
Figure 8-9 shows the electronic assembly for a
Hazardous Area Oxymitter 5000 with a
membrane keypad (upper view) and a
Hazardous Area Oxymitter 5000 with an LOI
(lower view).
Membrane Keypad
When Fault 7 is detected, The HEATER LED
flashes three times, pauses for three seconds,
and repeats.
1. If the case temperature exceeds 85°C
(185°F), the temperature control will shut
off and a fieldbus alarm will be sent.
2. This signifies that the environment where
the Hazardous Area Oxymitter 5000 is
installed exceeds the ambient temperature
requirements or that heat due to convec-
tion is causing case temperature to rise
above the limit.
3. Placing a spool piece between the stack
flange and the Hazardous Area Oxymitter
5000 flange may eliminate this problem.
4. If a spool piece does not solve the problem,
relocation is the only solution.
LOI
When Fault 7 is detected, the LOI displays the
"Board Temp Hi" message. Refer to the
comments in paragraphs 1 through 4 above.