EasyManuals Logo

Safran LRF 6019 User Manual

Safran LRF 6019
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
LRF 6019 Integrator Manual
914928_TML_LRF6019_en_Version C
Public: 04.2019
Confidential & Proprietary Safran Vectronix AG All rights reserved
Page: 55/70
6.6.6 Reset LRF module to default settings >LM,Rd*<CS><CR>
This command resets all LRF module’s configuration parameters to their default values.
An automatic restart of the LRF module is performed during the process.
Default parameters are:
Baud rate 57600
Range gate max 30000
Range gate min 00000
Timestamp deactivated
Reset LRF module settings command and response
Command
Inputs
Description
>LM,Rd*<CS><CR>
>LM,Rd*7B<0D>
Reset LRF module to default settings.
Example.
Response
Outputs
Description
No response expected.
Table 19: Reset LRF module settings command
Example of a reset LRF module to default settings command transaction:
Host LRF Module
Command frame: >LM,Rd*7B<0D>
LRF Module Host
Acknowledge frame: >AC*84<0D>
Prompt: <

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Safran LRF 6019 and is the answer not in the manual?

Safran LRF 6019 Specifications

General IconGeneral
BrandSafran
ModelLRF 6019
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals