Example: Enter 100 g as the upper limit of Class 1. Then enter 15%. When working
with 3 classes, this yields the following weight classes:
Class 0: up to the minimum load
Class 1: > minimum load - 100 g
Class 2: > 100 g – 115 g
Class 3: > 115 g - maximum load
When working with 5 classes, this yields the following weight classes:
Class 0: up to the minimum load
Class 1: > minimum load - 100 g
Class 2: > 100 g – 115 g
Class 3: > 115 g – 130 g
Class 4: > 130 g – 145 g
Class 5: > 145 g - maximum load
The values entered remain valid until deleted by pressing the c key or until
overwritten by a new value. They remain saved after the scale is switch-off.
Preparation t Open the APPLIC. / APPLIC.2 / CLASS. menu.
Available Parameter Settings * = Factory setting
MIN.INIT Minimum load for initialization 3.6
1 digit 1 digit 3.6.1*
2 digit 2 digits 3.6.2
5 digit 5 digits 3.6.3
10 digT. 10 digits 3.6.4
20 digT. 20 digits 3.6.5
50 digT. 50 digits 3.6.6
100 diG. 100 digits 3.6.7
200 diG. 200 digits 3.6.8
500 diG. 500 digits 3.6.9
1000 dI. 1000 digits 3.6.10
CTRL.SET Activate SET control output 4.3.
OUTPUT SET output 4.3.1*
OP.READY Ready to operate 4.3.2
OUTP.ACT Activation of outputs 4.7
OFF Off 4.7.1*
ALWAY.ON Always on 4.7.2
STABIL. On at stability 4.7.3
QTY. Number of classes 4.8.
3 CLASS 3 classes 4.8.1*
5 CLASS 5 classes 4.8.2
INPUT Parameter input 4.9.
WEIGHTS Weight values 4.9.1*
PERC.TAG Percentage values 4.9.2
PRINT Automatic printing 4.10.
manual Off 4.10.1*
automat On 4.10.2
) �t To save the setting, press the ) key.
( ( ... t To exit setup: Press the ( key several times.
Operation