Operating The PM295 55
AccS
PASS
1234
To change the user password:
Ä Press SELECT to choose the lower window.
Ä Adjust the password with the up/down arrow keys. The
password is up to four digits long.
Ä Press ENTER to return to the central window.
Store your password in a safe place. If you will not provide the correct
password, you will need to contact your local distributor for the superuser
password to override password protection.
Enabling/Disabling Password Checking
Ä From the central window, select the CtrL entry with the up/down arrow keys.
AccS
CtrL
OFF
To change the password protection mode:
Ä Press SELECT to choose the lower window.
Ä With the up/down arrow keys select OFF to disable
password protection, and select On to enable password
protection.
Ä Press ENTER to return to the central window.
3.4 Self-Test Diagnostics
The PM295 periodically performs self-test diagnostics. If the instrument fails the
self-test diagnostics, it discards the last measurement results, and an error code is
displayed for one second on all LEDs. Error codes are listed in Table 3.1. Frequent
failures may be result of excessive electrical noise in the region of the instrument. If
the instrument resets itself continuously, contact your local distributor.
Table 3-1 Self-Test Diagnostic Codes
Diagnostic Code Meaning
1 ROM error
2 RAM error
3 Watch dog timer reset
4 Sampling failure
5 Out of control trap
7 Timing failure
8 Power up