WARNING
• Make sure the analyzer is properly grounded.
• Before turning on the analyzer, make sure the input voltage meets the requirements.
CAUTION
• Using a pin-board may bring the electrical interference and the analysis results may be
unreliable. Please place the analyzer near the electrical outlet to avoid using the pinboard.
• Please use the original power cable shipped with the analyzer. Using other power cable
may damage the analyzer or cause unreliable analysis results.
4.2.3 General environment
• Optimal operating temperature: 50°F~86°F (10°C~30°C)
• Optimal operating humidity: 30%~85%
• Atmospheric pressure: 70 kPa~106 kPa
• The environment shall be as free as possible from dust, mechanical vibrations, loud
noises, and electrical interference.
• It is advisable to evaluate the electromagnetic environment prior to operation of this
analyzer.
• Do not use this analyzer in close proximity to sources of strong electromagnetic
radiation (e.g., unshielded intentional RF sources), as these may interfere with the
proper operation.
• Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close, such as a lab centrifuge.
• Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
• The environment must be ventilated.
• Install on a level surface.
• For indoor use only.