11 Troubleshooting the analyzer
11.1 Introduction
This chapter contains information that is helpful in locating and correcting problems that may
occur during operation of your analyzer.
NOTE:
This chapter is not a complete service manual and is limited to problems that are readily
diagnosed and/or corrected by the user of the analyzer.
11.2 Error information and handling
During the operation, if an error is detected, the analyzer will beep and display the
corresponding error message in the error information area at the bottom right of the screen.
Meanwhile, the indicator will turn red.
According to the severity of the errors, the colors of error messages are red, orange, blue and
green.
Red: Fatal error. When this kind of error occurs, the analyzer will stop running
immediately, and any further operation is prohibited.
Orange: Error that stops operation. When this kind of error occurs, the analyzer will stop
running immediately.
Blue: Error that restricts certain operations. When this kind of error occurs, the
analyzer can still continue with the current operation, but any other operations
related to the error will be restricted.
Green: Prompting error. When this kind of error occurs, the analyzer can still continue
with the current operation and other operations are not restricted.