Vers.: 20230710ENG
Page 67
J. EMC description
WARNING
Do not use this device in close proximity to sources of strong electromagnetic radiation (e.g.,
unshielded intentional RF sources), as these may interfere with the proper operation.
This equipment complies with the emission and immunity requirements of the EN61326–
1:2012 and EN61326–1:2013.
NOTE:
• It is the manufacturer's responsibility to provide equipment electromagnetic compatibility
information to the customer or user.
• It is the user's responsibility to ensure that a compatible electromagnetic environment for
the equipment can be maintained in order that the device will perform as intended.
K. Noise level
Maximal noise level: 80 dBA
NOTE:
Be sure to use and store the analyzer in the specified environment.
L. Operating environment
Optimal operating temperature: 50°F–86°F (10°C ~ 30°C)
Optimal operating humidity: 30% ~ 85%
Atmospheric pressure: 70 kPa ~ 106 kPa
M. Storage environment
Ambient temperature: 50°F–104°F (-10°C ~ 40°C)
Relative humidity: 10 ~ 90%
Atmospheric pressure: 50 kPa ~ 106 kPa
N. Running environment
Ambient temperature: 50°F–104°F (10°C ~ 40°C)
Relative humidity: 10% ~ 90%
Atmospheric pressure: 70 kPa ~ 106 kPa