Vers.: 20230710ENG
Page 35
2. Printing and transmission to LIS
If “Auto print after sample analysis” function is enabled, the analyzer will print reports
automatically; and if “Auto comm." function is enabled, the analysis results, sample and
patient information will be auto transmitted to LIS.
5.6 Shutdown
CAUTION
Do not start up the analyzer immediately after it is shut down. Wait for at least 10 seconds.
To ensure stable analyzer performance and accurate analysis results, be sure to perform the
“Shutdown” procedure to shut down the analyzer after it has been running continuously for
24 hours.
NOTE:
Perform the shutdown procedure to shut down the analyzer daily.
1. On the “Count” screen, tap MENU > SHUTDOWN.
2. Tap OK to perform the shutdown procedure.
3. When a dialog box pops up, asking you to perform Probe Cleanser maintenance,
present Probe Cleanser to the sample probe as instructed and start Probe Cleanser
maintenance. See Chapter 10: Servicing the analyzer to know how to do a Probe
Cleanser maintenance.
4. After Probe Cleanser maintenance is completed, the touch screen will black out and a
message will be displayed instructing you to power off the analyzer. Operate
accordingly.
5. Check if the waste container is full. If yes, empty the waste container and dispose of
the waste properly.
NOTE:
• Do not disconnect power during the shutdown process.
• If an error that will affect shutdown occurs during the showdown process, the analyzer will
resume its original status and report the error. See Chapter 11: Troubleshooting the
analyzer for solutions.