3. Remove the sample:
The sample probe will automatically aspirate the sample. When you hear the beep sound,
you may remove the sample.
4. Auto analysis and result reporting:
The analyzer will automatically run the sample. When the analysis is finished, the results,
scattergrams and histograms will be displayed on the screen.
NOTE:
During the analysis process, if errors like a clog or bubble occur, the analyzer will
automatically display the results of related parameters as invalid, and alarm information
will show on the error information area. See Chapter 11: Troubleshooting the analyzer
for the way to remove errors.
If the ambient temperature is outside the specified operating range, thus causing the
analyzer temperature (the temperature tested by the sensor inside the analyzer) to go
out its specified range, the analyzer will alarm for abnormal ambient temperature and the
analysis results may be unreliable. See Chapter 11: Troubleshooting the analyzer for
solutions.
5.5.3 Processing analysis results
1. Automatic saving of analysis results
This analyzer automatically saves sample results. When the maximum number of results
that can be saved has reached (40,000 records), the newest result will overwrite the oldest.